MS Sensor 3.0

Signal Calibration


The MS Sensor signal calibration feature allows an unlimited number of calibration levels to be defined for each signal. New calibration levels can be added by pressing the calibration table's "Add" button, and then entering the amount and signal response derived from the analysis of a suitable calibration standard. After entering all calibration levels for the signal, press the "Calculate" button, which will recalculate the calibration function based on the data in the calibration table:

For single-level calibrations, the calibration slope is determined from the amount and signal values entered in the calibration table, and the intercept will be set to zero. A linear regression is used to determine the slope and intercept for calibrations with 3 or more levels.

The calibration is also graphed on the "Calibration Plot" tab of the Signal- Calibration dialog box:

>>Back to MS Sensor Overview<<